Presentation is loading. Please wait.

Presentation is loading. Please wait.

ELECTRON MICROSCOPY Introduction Principles of operation of TEM

Similar presentations


Presentation on theme: "ELECTRON MICROSCOPY Introduction Principles of operation of TEM"— Presentation transcript:

1 ELECTRON MICROSCOPY Introduction Principles of operation of TEM
Sample preparation for TEM Advantages and disadvantages of TEM Principles of operation of SEM Sample preparation for SEM Advantages and disadvantages of SEM School of Biological Sciences Investigating micrographs

2 INTRODUCTION With the invention of the light microscope it was discovered that plant and animal tissues were made up of aggregates of individual cells. However, light microscopes are limited to approximately x1000 magnification and have poor resolution. Therefore not all the internal structures of a cell can be seen with a light microscope. In 1924 a French physicist by the name of de Broglie stated that a beam of electrons should behave in a similar way to a beam of light i.e with wave properties the wavelength should be shorter. Therefore an electron beam should give better resolution.

3 RESOLUTION When there is sufficient light, two points 0.2mm apart or more can be distinguished with the naked eye as being separate points. When this distance is less than 0.2mm, only one point is seen. This distance is called the resolving power (or resolution) of the eye. In other words resolution is the closeness two objects can be in proximity and still be perceived as two separate objects. Can’t see two separate objects Can see two separate objects

4 The invention of the electron gun
led to the development of the electron microscope. The metal tungsten filament is heated to about 2500oC which causes it to release electrons. Due to the large voltage difference between the filament and the anode plate the electrons are forced to flow in the direction of the arrow. The cathode shield increases the electron flow further and concentrates the electrons into a narrow beam. Back to principles of operation Filament Cathode High voltage generator Electron beam Anode

5 Two main types of electron microscopes;
TEM Transmission SEM Scanning TEM produces a high resolution image of the internal structures of cells. TEM uses the electrons that have passed through the specimen to form an image. SEM produces a three dimensional image of the specimen surface. A beam of electrons scans the whole specimen which then emits low energy, secondary electrons. This technique can be used to study whole cells.

6 When an electron beam strikes a specimen a number of events occur
When an electron beam strikes a specimen a number of events occur. Electrons are scattered depending on the nature of the material. If the electrons hit a dense array they are scattered out of the main beam and fewer electrons will reach the viewing screen. There is no fluorescence and that area appears dark. If the electrons pass a scarcity of atoms they travel straight through, hitting the viewing screen and causing fluorescence. That area will appear light. The image comes from the arrangement of light and dark patches on the screen.

7 PRINCIPLES OF OPERATION OF TEM
Electron gun Specimen holder Projection chamber

8 The main components of a TEM are:
THE ELECTRON GUN – produces an electron beam. THE COLUMN – uses electromagnetic lenses to control the beam and produce a magnified image IMAGE VIEWING AND RECORDING The image is produced on a fluorescent screen below which a shutter and camera are located. e e Condenser lens e e Specimen e Objective lens e e e Projector lens e e e e e e e e e e e e e e e Vacuum Fluorescent screen

9 Electrons only behave like light when they are manipulated in vacuum
Electrons only behave like light when they are manipulated in vacuum. Therefore the whole column is evacuated since atoms such as O2 and CO2 scatter the electrons. e e e e e e e

10 SAMPLE PREPARATION FOR TEM
The aims of sample preparation are as follows: To preserve the material in its natural state To ensure that the material withstands changes which might occur on exposure to atmosphere, vacuum and electron beam.

11 Fixation for TEM It is then placed into fixing solution
The tissue is cut into tiny pieces It is then placed into fixing solution

12 Dehydration and embedding of TEM
Tissue is placed in final embedding mixture and the resin is polymerised in the oven It is then placed in a dilute solution of resin embedding media Tissue is dehydrated in alcohol Specimen vials

13 Section cutting of TEM Sections are cut on an ultramicrotome with a glass or diamond knife. The sections are floated off the edge of the knife onto the surface of a water trough. The colour of the sections vary with thickness. When the sections are gold they are picked off the surface with a copper grid. The section on the copper grid is now ready for staining and viewing in the electron microscope. 3.05mm

14 Advantages Disadvantages
TEM Very good resolution Thin sections are effectively two dimensional slices of tissue and do not convey the three dimensional arrangement of cellular components magnification Artefacts may be created Can see sub-cellular components and measure them

15 PRINCIPLES OF OPERATION OF SEM
Electron gun Detector Detector Image viewing Specimen chamber Control panel

16 SEM uses electrons that are emitted from the specimen surface.
The specimen is scanned with a very fine beam of electrons. These are scattered as they hit high and low points in the specimen. The scattered electrons are measured by a detector and used to control a second beam which forms an image on a TV screen Electron source T.V. Monitor e Electron beam e e e e e eee Specimen Detector Vacuum

17 Sample preparation for SEM
The preservation used will usually determine which drying process to use. There are two basic methods of drying the specimen: Freeze-drying – used after freezing Critical point drying – used after chemical fixation and dehydration

18 Freeze drying for SEM Sample placed in nitrogen slush to maintain it’s structure Then placed in liquid nitrogen to allow easy handling Sample placed in copper holder and………. …..placed in freeze drier Sample is mounted on a stub

19 Critical point drying for SEM
Sample is chemically fixed Then dehydrated with alcohol Sample is placed in critical point drier. Here the sample is flushed several times with liquid CO2. The pressure and temperature is then raised which converts the liquid CO2 to gas. The gas is then vented off slowly. The sample is removed and mounted on a stub

20 Coating the specimen for SEM
Most biological specimens are poor conductors and poor emitters of secondary electrons therefore the surface of the sample needs to be coated with a thin layer of a conducting material. There are two ways to do this: Sputter coating Evaporation of carbon

21 Sputter coating for SEM
When power passes to the anode, the noble metal evaporates (called the plasma effect) and the metal falls onto and coats the specimen. ANODE METAL A sputter coater CATHODE

22 Evaporation of carbon for SEM
Two carbon rods are placed end to end. One of the rods is sharpened to a point. These are placed in a vacuum and the specimen is placed below them. When electricity passes through the carbon rods, the carbon tip evaporates and the carbon falls onto and coats the specimen. CARBON RODS CARBON RODS

23 Advantages Disadvantages
SEM Provides great depth of focus Micrographs show a 3D image of specimen Smaller and simpler in comparison to TEM Only surface features seen Resolution attainable is not very high (approx 10nmn)

24 Investigating Micrographs
Transmission Electron Micrographs Scanning Electron Micrographs Can you spot the differences between the two types of electron microscopy?

25 Transmission Electron Micrographs
Virus particles Plasma membrane Endoplasmic reticulum Golgi membranes Mitochondrion Section of mammalian cell

26 Scanning Electron Micrographs
Sample of geranium petal showing the cone shaped projections and the internal structure.

27 Measuring Micrographs
How to work out the size of an organelle? Measured size Magnification Magnification of micrograph is X100000 Measured size = 80mm Convert to m = 80000  m 100000 = 0.8m or 800nm

28 for using this programme. We hope that it has been useful!
This programme was developed as part of a work placement project by Sumerah Khan and Sheerin Dariani THANKS TO: Chris Gilpin, Ian Miller Les Lockey, Samantha Newby

29 References B. Schotanus (1980) Electron microscopy, what is it ? Marketing electron optics. Philips Export B.V. Eindhoven. Dr Yvonne Miller (1998) Preparation of specimens for TEM and SEM. Mike Mahon, Chris Gilipin, Ian Miller (2000) Microscopy and analysis University of Manchester - School of Biological Sciences. Sam Newby (2000) Freeze drying and critical point drying EMPGU. 5) Specimen preparation (1991) (21/1/00) world/SEM/ specimenprep.html pages 1-2. 6) Dr. Ron Butler (1980) Transmission electron microscopy, What an SEM is ?, Aims of specimen preparation and Electron microscopy unit. EMPGU

30 1. What is the resolving power of the naked eye?
Questions 1 & 2 1. What is the resolving power of the naked eye? A 200 m B 0.02m C 0.002m D 2mm 2. The filament of a transmission electron microscope is made up of which element? A copper B tungsten C carbon D gold

31 True or false? true or false ?
Questions 3 & 4 3. Electrons flow away from the filament because of the large voltage difference between the filament and the anode plate True or false? 4. In electron microscopy, the lenses used to magnify the image are made of electromagnets true or false ?

32 6. A vacuum is needed in the electron microscope to
Questions 5 & 6 5. Which of the following is the first step in the processing of biological material for transmission electron microscopy? A Dehydration B Sectioning C Fixation D Embedding 6. A vacuum is needed in the electron microscope to A Pull the electrons onto the specimen B Eliminate molecules of nitrogen, oxygen or carbon dioxide C Pull the specimen into the column D Prevent secondary radiation affecting the microscope control panel

33 Questions 7 & 8 7. The difference between TEM and SEM is that in TEM, secondary low energy electrons are used to produce an image. True or false 8. Which of the following structures could not be seen with a light microscope but could be seen with a transmission electron microscope? A Nucleus B Cell wall C Ribosome D Golgi apparatus

34 9. Which of the following is not an advantage of TEM?
Questions 9 & 10 9. Which of the following is not an advantage of TEM? A High resolution B High magnification C Three dimensional detail D Detail of sub cellular components 10. Which of the following statements about SEM is not true? A The specimen is usually coated with gold B Resolution is excellent C Samples for SEM can be chemically fixed or freeze dried D Thin sections of tissue are not necessary

35 wrong please choose again
Try again wrong please choose again

36 Answer question 1 Answer: A this is the same as the value quoted in the text i.e. 0.2mm

37 Answer question 2 Answer: B a tungsten filament is used

38 Answer correct Correct well done

39 Answer question 7 Answer: this is false as secondary electrons are used to form an image in SEM

40 Answer question 5 Answer: the material has to be fixed to prevent distortion and decomposition

41  Answer: B molecules in the air would cause scattering of electrons
Answer question 6 Answer: B molecules in the air would cause scattering of electrons

42 Answer question 8 Answer: C it is possible to see the other structures with a light microscope

43 Answer question 9 Answer: C TEM gives a two dimensional image

44  Answer: B SEM has relatively poor resolution compared with TEM
Answer question 10 Answer: B SEM has relatively poor resolution compared with TEM

45 Scanning Electron Micrographs
Sample of geranium petal showing the cone shaped projections and the internal structure.

46 Transmission Electron Micrographs
Virus particles Plasma membrane Endoplasmic reticulum Golgi membranes Mitochondrion Section of mammalian cell

47 Q 11What is the size of the virus
Questions 11 & 12 Q 11What is the size of the virus Q12 What is the size of the mitochondrion Magnification X100000 Magnification X13000 Measured size = 9mm Measured size = 11mm A 90mm A 846mm B 90nm B 8.46nm C 90m C 8.46m D 900nm D 846nm


Download ppt "ELECTRON MICROSCOPY Introduction Principles of operation of TEM"

Similar presentations


Ads by Google