Presentation is loading. Please wait.

Presentation is loading. Please wait.

Integrating the Healthcare Enterprise (IHE) Patient Care Devices Domain (PCD) Alarm Communication Management (ACM) Requirements for AM – AC Interoperability.

Similar presentations


Presentation on theme: "Integrating the Healthcare Enterprise (IHE) Patient Care Devices Domain (PCD) Alarm Communication Management (ACM) Requirements for AM – AC Interoperability."— Presentation transcript:

1 Integrating the Healthcare Enterprise (IHE) Patient Care Devices Domain (PCD) Alarm Communication Management (ACM) Requirements for AM – AC Interoperability Monroe Pattillo Philips Patient Monitoring Informatics (PMI) Boca Raton, FL, USA IHE Patient Care Devices Domain ACM Working Group © 2009 ACCE, HIMSS, IHE

2 2 This presentation can be found at... ftp.ihe.net/Patient_Care_Devices/Presentations/ACM/ ACM AM-AC Requirements 01.ppt

3 3Outline ACM AM – AC MessagesACM AM – AC Messages AM-AC Protocol RequirementsAM-AC Protocol Requirements What are some possibilities?What are some possibilities? What do we do next?What do we do next?

4 4 Alarm Communicator AC Alarm Communicator AC Alarm Manager AM Alarm Manager AM Alarm Reporter AR ACM AM – AC Messages Alarm Archiver AA Report Alarm PCD-04 → ← PCD-05 Report Alarm Status Disseminate Alarm PCD-06 → ← PCD-07 Report Dissemination Alarm Status PCD-08 ↑ Subscribe To Alarm PCD-04 ↓ Report Alarm PCD-04, PCD-05, PCD-08 all HL7 v2.6 → v2.7 PCD-06, PCD-07 Data items defined Currently using SMTP as open standards based transport, but needs better

5 5 PCD-04 Message An HL7 ORU^R01 message containing segments for... MSHHeader PIDPatient Identification PV1Location Information OBRUnique Observation Identification ORCOptional Order Information OBXObservations, containing Type, Units, Range, Text description, Type, Units, Range, Text description, Priority (indication - low, med, high), Priority (indication - low, med, high), Type (physiological, technical, etc.), Type (physiological, technical, etc.), State (onset, continue, escalate, end) State (onset, continue, escalate, end) NTENotes and Comments

6 6 PCD-06 AM to AC Message AM to AC PCD-06 – Disseminate Alarm Sends the message to the AC’s device by way of AC Identification of message recipients Message, containing Location associated with alarm Patient associated with alarm Message text (incl. physiological info.) Identifier unique identifier Callback (telephony dial back) Contextual Data Reference (URL) Comment Evidentiary Data

7 7 PCD-07 AC to AM Message AC to AM PCD-07 – Report Dissemination Alarm Status Sends message deliver status updates to the AM Accepted by AC (delivered to carrier) Undeliverable to endpoint Delivered to endpoint Accepted by endpoint Accepted by endpoint as true positive (info) Accepted by endpoint as true positive however not clinically relevant (operator note) Accepted by endpoint as false positive (info) Rejected by endpoint Canceled by endpoint (escalation, not alarm)

8 8 AM-AC Protocol Requirements Must be an existing protocol IHE uses standards, it does not develop them Must be an open protocol Not proprietary Must have unrestricted use

9 9 AM-AC Protocol Requirements Preferences/Focus Contemporary, Healthcare, People Communications XML based – extensible HTTP transport – works with firewalls and NATs Multiple Choice Response (MCR) – “soft keys” Pushed, not polled status updates – quicker Presence/Status – for escalation responsiveness Dynamic – online/offline, busy, BRB, away, DND TLS capable – SSL securable, messages contain ePHI WCM Supportive – hyperlink or large evidentiary data

10 10 AM-AC Protocol Requirements ACM Specific Messaging Items Device is on network identified by ID/Number/Name Delivery target is ultimately people, not systems Dissemination/Status unsolicited, not subscribed PCD-04 round trip event message identification Many status potentials in status enumeration delivery/not at multiple levels (carrier, device) human interactions read receipt multiple response possibilities more than accept/reject or timeout

11 11Resources This presentation can be found at... ftp.ihe.net/Patient_Care_Devices/Presentations/ACM/ ACM AM-AC Requirements 01.ppt

12 12 IHE PCD Contacts PCD Co-Chairs Todd Todd Ken Ken PCD Secretariat Manny Manny ACM WG Co-Chairs Monroe Monroe John John

13 13 Thank You