Presentation is loading. Please wait.

Presentation is loading. Please wait.

Pacemaker Patient Follow-up Module 8

Similar presentations


Presentation on theme: "Pacemaker Patient Follow-up Module 8"— Presentation transcript:

1 Pacemaker Patient Follow-up Module 8
Student Notes This is Module 8 in the CorePace series and covers the basics of performing pacemaker follow-up. This module will discuss the baseline information necessary for working toward more advanced knowledge in pacemaker operation. It is possible that you may require additional supplemental materials to enhance your knowledge, or provide more practice. If you feel this is necessary for you, ask your instructor for suggestions on books or other tools. Instructor Notes This module should take approximately 2 hours to cover. While delivering the module, engage the learners by asking questions and getting them to talk based on their previous knowledge. Evaluate the learners by delivering the knowledge check at the end of this module. An acceptable score is 90%. World Headquarters Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN USA Internet: Tel: (763) Europe Medtronic International Trading Sàrl Route du Molliau Ch Tolochenaz Switzerland Tel: (41 21) Asia-Pacific Medtronic International, Ltd. 16/F Manulife Plaza The Lee Gardens, 33 Hysan Avenue Causeway Bay Hong Kong Tel: (852) Canada Medtronic of Canada Ltd. 6733 Kitimat Road Mississauga, Ontario L5N 1W3 Tel: (905) Toll-free: 1 (800) Medtronic USA, Inc. Toll-free: 1 (800) (24-hour technical support for physicians and medical professionals) Latin America Medtronic USA, Inc. Doral Corporate Center II 3750 NW 87th Avenue Suite 700 Miami, FL 33178 USA Tel: (305) UC d EN Medtronic, Inc. Minneapolis, MN January 2008

2 Objectives List the steps for performing a pacemaker follow-up
Locate threshold, sensing and impedance data on a Quick Look™ screen Identify broad trends in Cardiac Compass® and HF Management Reports Identify additional resources or information useful for performing patient follow-up Student Notes Instructor Notes

3 Pacemaker Patient Follow-up
Routine device follow-up Typically 1 per year for chronic single chamber and 2 per year for chronic dual chamber pacemakers These guidelines evolved because the primary purpose of the follow-up was to evaluate the device Increasing sophistication of follow-up tools and the diagnostics the devices provide, may be changing the purpose of the follow-up, but have not affected the guidelines Student Notes Instructor Notes

4 Pacemaker Patient Follow-up Steps
Evaluating the device Determine the battery voltage Check the lead impedance Test capture thresholds Test sensing thresholds Perform a magnet/non-magnet test (required by Medicare in some US states) Student Notes Note: The order you execute these steps is irrelevant – just be sure all steps are executed and the data recorded per the clinic protocol. Instructor Notes

5 Pacemaker Patient Follow-up Evaluating the Device
Battery voltage To verify the pacemaker’s ability to operate and estimate the remaining longevity Lead impedance To verify that the leads are electrically intact Capture thresholds To verify an appropriate pacing safety margin Sensing thresholds To verify an appropriate sensing safety margin To observe the underlying rhythm Magnet/non-magnet test To verify Recommended Replacement Time (RRT) has not been reached Student Notes Recommended replacement time is often abbreviated as RRT. Some older devices refer to this same time as early replacement indicator (ERI). Instructor Notes Ask: Why is battery voltage evaluated? To verify the pacemakers ability to operate and estimate the remaining longevity Ask: Why is lead impedance evaluated? To verify that the leads are electrically intact Ask: Why are capture thresholds evaluated? To verify an appropriate pacing safety margin Ask: Why are sensing thresholds evaluated? To verify an appropriate sensing safety margin To observe the underlying rhythm Ask: Why is a magnet/non-magnet test evaluated? To verify the recommended replacement time (RRT) has not been reached

6 Pacemaker Evolution for the Follow-up Clinic
Battery voltage Estimate of remaining time to replacement Most modern devices provide this estimate in some fashion Most modern devices will also indicate impending elective or recommended replacement Lead impedance Measured with interrogation Long-term trends, if available, are very useful Pacing thresholds In some devices this test is automatic and performed routinely (e.g., daily) Sensing thresholds In some devices this test is automatic and performed routinely (e.g., every intrinsic event) Magnet test Formerly only way to determine if Recommended Replacement Time (RRT) had been reached Less of a need as devices now routinely report RRT condition Student Notes Instructor Notes

7 Battery Voltage Estimating Remaining Longevity
Max voltage in a pacemaker battery is typically 2.8 V RRT at 2.5 Volts Voltage information is provided on the programmer Many devices estimate remaining longevity based on % pacing, lead impedances, and pacing outputs Magnet Test evolved as a way to evaluate if RRT criteria is met via a simple telephone or in-clinic check When a magnet is applied, verify if the pacemaker shows normal magnet behavior or RRT magnet behavior For some pacemakers, a magnet initiates a Threshold Margin Test (TMT) Student Notes The Threshold Margin Test assists with verifying capture thresholds. The TMT may indicate that loss-of-capture is possible but cannot verify that the safety margin is adequate. For Medtronic pacemakers that have TMT: A dual chamber pacemaker delivers three asynchronous AV sequential pulses at a rate of 100 ppm, with a paced AV interval of 100 ms The first two sequences of pulses are delivered at the programmed Amplitude The third sequence is delivered at a 20% reduction of the programmed Amplitude At the completion of the TMT, pacing is forced to a rate of 85 bpm in the magnet mode Instructor Notes Ask: Where can you find the magnet rates for various pacemakers? In the Pacemaker and ICD Encyclopedia

8 Resource Pacemaker And ICD Encyclopedia
All device manufactures share essential device information All publish encyclopedias Medtronic’s version is available on-line, as a download for both Palm and Pocket PC All include information on leads, devices, model numbers, x-ray identification, etc. Includes Beginning-of-Life (BOL) magnet mode and rates RRT magnet mode and rates Other RRT indicators (e.g., pulse width “stretching”) Student Notes Note: Magnet rates for most new models of Medtronic pacemakers are: BOL: 85 ppm RRT: 65 ppm Instructor Notes

9 Status Check Determine the magnet and non-magnet modes, and rates for a Kappa® 700, Model KDR731. Click for answer Student Notes Many device manufactures maintain Technical Service Departments, which are usually an excellent resource: Medtronic’s phone: (U.S.) Instructor Notes Distribute the Pacemaker and ICD Encyclopedias (can be found on Medtronic Connect website) Ask: Find the magnet rate and mode for a Kappa 700, Model KDR731, at Beginning-of-Life (BOL). Magnet rate: 85 ppm Mode: DOO Ask: Find the magnet rate and mode for a Kappa 700, Model KDR731, at Early Replacement Indicator (ERI). Magnet rate: 65 ppm Mode: VOO

10 Status Check Find the Estimated Longevity
Click for answer Battery Voltage can be found by selecting “Battery and Lead Measurements” under the Data icon. Student Notes Instructor Notes

11 Magnet Testing If required:
Use a donut magnet (or the programming head) Record the ECG with and without the magnet Medicare guidelines call for a 30-second recording of each Observe the ECG for asynchronous pacing with the magnet Measure the pacing rate Some modern devices provide a specific magnet test as an option on their programmer Student Notes Instructor Notes

12 Lead Impedance Can you recall why knowing the lead impedance is important? Because the lead impedance can warn you of a lead insulation failure or a lead conductor fracture At implant, it could indicate a loose set screw Can you recall the expected range of lead impedances? Normally, Ω, although some specially designed leads may be higher Click for answer Student Notes Instructor Notes Click for answer

13 Status Check Find the lead impedances
Click for answer These trends show stable impedances. For more information, touch the chevron. Student Notes Instructor Notes

14 Pacing Thresholds Can you recall the pacing output safety margin on a chronic system? Multiply the amplitude threshold by two Can you recall the maximum you would want to program the amplitude, assuming a threshold of < 1.0 V? Normally, try to keep the output amplitude at < 2.5 V Patient safety is paramount Click for answer Click for answer Student Notes Two of the ways to maintain an adequate pacing safety margin for chronic leads are: While holding the pulse width constant, multiply the amplitude threshold by two, or While holding the amplitude constant, multiply the pulse width by three Instructor Notes Discuss the following example with the participants: Threshold testing results in 1.5 V at 0.2 ms Battery voltage is 2.6 V Chronic leads The clinic has established a minimum pacing amplitude of 2.0 V Ask: What would you program the outputs to? 2.0 V at 0.6 ms Explanation: Multiplying the amplitude by two and achieving an output of 3.0 V would be an appropriate safety margin. However, the voltage output exceeds the battery voltage. Multiplying the pulse width by three and holding the voltage at 1.5 would also be an adequate safety margin. However, the clinic established minimum output of 2.0 V would be inappropriate. Therefore, an output of 2.0 V at 0.6 ms maintains an adequate safety margin, while keeping the voltage above the clinic established minimum. Another option would be to retest the threshold at a higher pulse width.

15 Status Check Find the pacing thresholds
Click for answer These trends show stable thresholds. Note how the atrial threshold has decreased since implant. For more information, touch the chevron. Student Notes Instructor Notes

16 Threshold Testing Manual Methods
The device automatically performs threshold tests But where would you find a manual threshold (or other) test? Student Notes Some different methods for performing threshold tests include: Capture Management Medtronic’s automatic atrial and ventricular capture verification algorithm Runs automatically at clinician determined intervals Automatically reprograms the pacemaker to 2X (or other) safety margin Strength Duration Tests the amplitude at a fixed pulse width Then, tests pulse width at twice the threshold amplitude The clinician determines capture/loss of capture Useful in guiding programming decisions Auto decrement Test either the amplitude or the pulse width One value is held constant, the other is tested Fast test, useful when system is chronic and thresholds stable Instructor Notes Click for answer

17 Status Check Recall the definition of the pacing threshold…
The minimum output at which the myocardium is consistently captured, outside of its refractory period. Click for answer Student Notes Instructor Notes

18 Performing a Manual Pacing Threshold Test
Process Tell the patient what you are going to do Force pacing Periodically lower the test value (amplitude or pulse width) Stop the test once Loss-of-Capture (LOC) is seen on the ECG The value just above LOC is the threshold 0.75 V 1.25 V 1.00 V Student Notes There are many acceptable ways to force pacing. Here are few for the given chamber: Atrial: Using AAI on a patient with AV conduction, increase the pacing rate to above the intrinsic rate Atrial: Using DDD on a patient without AV conduction, increase the pacing rate to above the intrinsic rate Ventricular: Using VVI, increase the pacing rate to above the intrinsic rate Ventricular: Using DDD, decrease the AV to below the intrinsic AV interval Periodically lowering the test value is automatically accomplished if you utilize the Auto Decrement feature on the Threshold Test screen. Instructor Notes Ask: What is the threshold? 1.25 V Set up a VIP-II and demo device with varying degrees of inappropriate outputs (i.e., an atrial pacing output of ms, with an atrial threshold of ms). Have each student perform an atrial and ventricular threshold test. Make sure that each participant can appropriately: Identify loss of atrial and ventricular capture Label the capture thresholds Recommend appropriate programming changes to pacing outputs, if necessary

19 Status Check Find the P- and R-wave measurements
Click for answer To see the P- and R-wave trends, touch the chevron. Student Notes Instructor Notes

20 Sensing Tests Manual Methods
This device automatically measures P- and R-waves But where would you find a manual sensing (or other) test? Student Notes Instructor Notes Click for answer

21 Status Check What is the definition of the sensing threshold?
The minimum signal size required to inhibit (or trigger) the pacemaker Click for answer Student Notes Instructor Notes

22 Performing a Manual Sensing Test
Process Tell the patient what you are going to do Reduce (if necessary) the pacing rate to allow the intrinsic events to occur Periodically increase the test value (e.g., for 0.5 mV to 1.0 mV) Observe the ECG for loss of inhibition (pacing despite the presence of P- or R-waves) The value, just lower than when pacing occurs, is the size of the P- or R-wave Student Notes The periodic increase of the test value is automatic for most current pacemakers. Instructor Notes Set up a VIP-II and demo device with varying degrees of inappropriate sensitivity settings (i.e., an R-wave of 3 mV, with a ventricular sensitivity setting of 2.5 mV). Have each student perform a P- and R-wave sensing test. Make sure that each participant can appropriately: Identify loss of atrial and ventricular sensing Recommend appropriate programming changes to sensitivity, if necessary

23 Routine Testing With almost all manufacturers, routine threshold and sensing tests can be performed on a programmer interface that guides you through the steps Observe the patient carefully for symptoms or complaints Follow the on-screen directions and closely observe the ECG for changes Student Notes Instructor Notes

24 Other Routine Tests Underlying rhythm Retrograde conduction Etc.
Can frequently ascertain the underlying rhythm during the sensing test Optionally, you can perform this test with the programmer Retrograde conduction Etc. Done per clinic protocol, or only if needed Student Notes Instructor Notes

25 So We’ve Checked the Pacemaker
Is that all there is to it? Well… Is Rate Response programmed appropriately? Can the patient achieve rates to support activities of daily living? Is the patient having any arrhythmias? What kind of arrhythmias, what rates, what triggers? Can we program the device to extend its longevity? Can we encourage intrinsic events? …are just some of the questions we can answer by looking at the device diagnostics. Student Notes Activities of daily living are often abbreviated as ADL. Accurate diagnostics depend on: The battery is within the normal operating range The leads are electrically intact When the pacemaker paces – it captures All intrinsic cardiac events are sensed and no others Instructor Notes For each question: Ask: What other questions can be asked to uncover more information?

26 Optimizing Pacemakers for Patients
Always evaluate the rate histograms Look for a “stair-case” distribution of rates Ask about the patient’s ability to achieve their desired levels of activity Student Notes LRL = Lower Rate Limit UTR = Upper Tracking Rate USR = Upper Sensor Rate Instructor Notes Rates occurring below the LRL are explained in “CorePace Module 9: Troubleshooting,” in case 9. Ask: What do you think about this distribution of rates? This distribution is ok, but it depends on the patient’s symptoms and desired activity level Rates below the LRL may occur because of a timing anomaly. Rates above the UTR/USR may indicate an arrhythmia.

27 Optimizing Pacemakers for Patients
Always evaluate for the presence of arrhythmia What type? Is the patient symptomatic? Student Notes Instructor Notes Ask: Why are these important? Information on atrial arrhythmias is important to physicians, so they can manage co-morbidities, such as risk for stroke Cardiac Compass is a simple way to stay informed of trends in the patient’s atrial arrhythmia burden. Cardiac Compass also provides trends on the ventricular response to these arrhythmia.

28 Episode Specific Diagnostics
In pacemakers, stored EGMs can be useful: Confirming accuracy of other arrhythmia diagnostics For example – is oversensing triggering arrhythmia collection? Collecting arrhythmia triggers An EGM is an ECG recorded via the pacemaker’s leads and stored in the device Student Notes Instructor Notes Stored EGM of an episode of AF

29 Optimizing Pacemakers for Patients
Always evaluate the percent pacing Many devices have a percent pacing counter, or look to the histograms for this information Device longevity is affected by pacing outputs High outputs decrease longevity Low outputs (below about 2.0 V) don’t have as dramatic of an affect So reducing the percentage of unnecessary pacing will improve device longevity and there may be other benefits to patients (more later) Student Notes Instructor Notes

30 Reducing the Pacing Percentage
Managing AV delays Use auto AV extension algorithms (e.g., Medtronic’s Search AV+) with Auto-PVARP options to: Reduce unnecessary right ventricular pacing Allow higher UTR Guard against retrograde conduction Student Notes Instructor Notes Ask: In patients with a CRT device – designed to restore ventricular synchrony via bi-ventricular pacing – what percent ventricular pacing is appropriate? The goal for CRT is 100% ventricular pacing

31 Reducing the Pacing Percentage
Search AV+ example: AV intervals are scanned If the majority end in pacing (8/16) the AV delay is automatically increased Student Notes Instructor Notes Note: AV scanning algorithms will typically still result in ventricular pacing about 40-50% of the time.

32 Using Other Diagnostics
Some devices include indicators of HF status and can be very useful For example: HR variability, Average day/night rates, Activity trends Otherwise be guided by: The patient’s diagnosis, complaints, symptoms Student Notes Instructor Notes

33 Clinic Evolution Taking Advantage of Pacemaker Automaticity
Devices do not routinely require extensive testing to confirm they are operating correctly But when troubleshooting a particular problem is required, most devices today offer detailed data that may help assist with analysis of device, lead and disease progression concerns If all tests can be performed automatically, and if we could view all device diagnostics remotely, why not: Use the pacemaker clinic only when a device has a problem or requires reprogramming Allow the physician to determine if routine clinic visits are warranted? Student Notes The point is not to avoid patients – the point is to use your limited resources most efficiently and to the most benefit. Remote monitoring, e.g., the Medtronic CareLink®, permits patients to download to a secure site, which the clinic can visit at their convenience. This may eventually evolve to wireless downloads and automatic alerts, requiring no patient initiation. What would you do with the time saved? How could you use this time to better manage all of your patients? Instructor Notes

34 Medtronic CareLink® Network Example
Clinic A ****** Click to Login Student Notes Security measures, such as full username and password authentication, ensure data privacy and integrity. Instructor Notes

35 CareLink™ Network Example
Pick the patient transmission Student Notes Here we click on the “Last Sent” of the InSync Sentry® patient, to view his reports. Instructor Notes Note: the patient information is fictitious

36 CareLink™ Network Example
Student Notes The CareLink™ report for a pacemaker is a pdf file. This slide shows the top half of the first page of the report. Instructor Notes

37 Status Check Case 1 This patient has had a DDDR pacemaker for 6 years
Functioning normally On a routine telephone check, the following strip is transmitted The patient’s magnet is in place What conclusions can you draw? Click for answer Asynchronous pacing Student Notes Instructor Notes A-A interval 705ms V-V interval 705ms Normal magnet behavior, operating in DOO mode at 85 ppm

38 Status Check Case 2 This patient was implanted about 6 months ago for SND He came to the clinic for a wound check 2 weeks post op. This is first IPG clinic visit. This (hypothetical) pacemaker requires that you do all the testing. How is the device currently programmed? Student Notes Instructor Notes Click for answer

39 Status Check Case 2 Results of pacemaker testing:
Battery voltage: 2.78 V Sensing: P-waves mV R-waves >22.8 mV Underlying rhythm Sinus Brady at 50 bpm Thresholds Atrial 1.0 V tested at 0.4 ms Ventricular 0.8 V tested at 0.4 ms Student Notes Instructor Notes

40 Status Check Case 2 What programming changes (if any) would you recommend? Re-program the A and V outputs to 2.0 V at 0.4 ms is reasonable. The leads should be stable at 6 months post implant. Would you run any other tests or investigate any further? The underlying rhythm indicates intact conduction. Consider programming a Search AV algorithm to reduce unncessary ventricular pacing. Check the rate histograms and question the patient for the adequacy of his rate response. Check other diagnostics for the presence of arrhythmia. Click for answer Click for answer Student Notes Instructor Notes

41 Status Check Case 3 This elderly patient had a DDDR implanted for sinus node dysfunction 8 weeks ago Discovered sinus node disease with pauses on routine physical Past medical history includes hypertension She denies palpitations Treatment: Hydrochlorothiazide (blood pressure medication) Patient comes to the clinic for her first routine check All values within normal and expected limits Atrial and Ventricular outputs reduced to a 2.5x safety margin No other changes made Student Notes Instructor Notes

42 Status Check Case 3 Cardiac Compass Report Is she having AF?
How much AF is she having? Student Notes Instructor Notes Click for answer Episodes totaling 8 hours on one day, about 4 hours on two days, and several shorter episodes 1 min to 1 hour durations thereafter.

43 Status Check Case 3 Cardiac Compass Report
Is the ventricular response to AF well controlled? Click for answer Student Notes Instructor Notes Yes, as a very low percent of the ventricular response to AF is > 100 bpm.

44 A Couple of Questions In patients with paroxysmal AF (PAF), how likely are they to experience symptoms of AF? What is the risk of stroke or CVA to someone with untreated AF? Click for answer Asymptomatic PAF occurs 12.1 times as often as symptomatic PAF in symptomatic patients [Page et al, Circulation 89(1):224-7, 1994] Asymptomatic PAF occurs in 22-27% of patients with clinical improvement [ Wolk et al, Int J Cardiol 54: , 1996] Student Notes Instructor Notes Click for answer 5-7% per year. The take-away: AF is silent and dangerous.

45 Status Check Case 3 Based on the Cardiac Compass® data, the MD decides to add Sotalol 160 bid and Coumadin The patient returns 6 months later Is her arrhythmia status improved? Click for answer Student Notes Instructor Notes Yes, it has improved. Her last episode was at the end of May.

46 A Couple More Questions
For this patient, how would you make the diagnosis of PAF without these kinds of diagnostics? She had no symptoms She was not in AF on clinic visits For this patient, how would you evaluate the effectiveness of your treatment without these kind of diagnostics? Her Medication regimen was not that effective in suppressing her AF Student Notes Instructor Notes

47 Brief Statements Indications
Implantable Pulse Generators (IPGs) are indicated for rate adaptive pacing in patients who ay benefit from increased pacing rates concurrent with increases in activity and increases in activity and/or minute ventilation. Pacemakers are also indicated for dual chamber and atrial tracking modes in patients who may benefit from maintenance of AV synchrony. Dual chamber modes are specifically indicated for treatment of conduction disorders that require restoration of both rate and AV synchrony, which include various degrees of AV block to maintain the atrial contribution to cardiac output and VVI intolerance (e.g. pacemaker syndrome) in the presence of persistent sinus rhythm. Implantable cardioverter defibrillators (ICDs) are indicated for ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias. Cardiac Resynchronization Therapy (CRT) ICDs are indicated for ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias and for the reduction of the symptoms of moderate to severe heart failure (NYHA Functional Class III or IV) in those patients who remain symptomatic despite stable, optimal medical therapy and have a left ventricular ejection fraction less than or equal to 35% and a QRS duration of ≥130 ms. CRT IPGs are indicated for the reduction of the symptoms of moderate to severe heart failure (NYHA Functional Class III or IV) in those patients who remain symptomatic despite stable, optimal medical therapy, and have a left ventricular ejection fraction less than or equal to 35% and a QRS duration of ≥130 ms. Contraindications IPGs and CRT IPGs are contraindicated for dual chamber atrial pacing in patients with chronic refractory atrial tachyarrhythmias; asynchronous pacing in the presence (or likelihood) of competitive paced and intrinsic rhythms; unipolar pacing for patients with an implanted cardioverter defibrillator because it may cause unwanted delivery or inhibition of ICD therapy; and certain IPGs are contraindicated for use with epicardial leads and with abdominal implantation. ICDs and CRT ICDs are contraindicated in patients whose ventricular tachyarrhythmias may have transient or reversible causes, patients with incessant VT or VF, and for patients who have a unipolar pacemaker. ICDs are also contraindicated for patients whose primary disorder is bradyarrhythmia.

48 Brief Statements (continued)
Warnings/Precautions Changes in a patient’s disease and/or medications may alter the efficacy of the device’s programmed parameters. Patients should avoid sources of magnetic and electromagnetic radiation to avoid possible underdetection, inappropriate sensing and/or therapy delivery, tissue damage, induction of an arrhythmia, device electrical reset or device damage. Do not place transthoracic defibrillation paddles directly over the device. Additionally, for CRT ICDs and CRT IPGs, certain programming and device operations may not provide cardiac resynchronization. Also for CRT IPGs, Elective Replacement Indicator (ERI) results in the device switching to VVI pacing at 65 ppm. In this mode, patients may experience loss of cardiac resynchronization therapy and / or loss of AV synchrony. For this reason, the device should be replaced prior to ERI being set. Potential complications Potential complications include, but are not limited to, rejection phenomena, erosion through the skin, muscle or nerve stimulation, oversensing, failure to detect and/or terminate arrhythmia episodes, and surgical complications such as hematoma, infection, inflammation, and thrombosis. An additional complication for ICDs and CRT ICDs is the acceleration of ventricular tachycardia. See the device manual for detailed information regarding the implant procedure, indications, contraindications, warnings, precautions, and potential complications/adverse events. For further information, please call Medtronic at and/or consult Medtronic’s website at Caution: Federal law (USA) restricts these devices to sale by or on the order of a physician.

49 Brief Statement: Medtronic Leads
Indications Medtronic leads are used as part of a cardiac rhythm disease management system. Leads are intended for pacing and sensing and/or defibrillation. Defibrillation leads have application for patients for whom implantable cardioverter defibrillation is indicated Contraindications Medtronic leads are contraindicated for the following: ventricular use in patients with tricuspid valvular disease or a tricuspid mechanical heart valve. patients for whom a single dose of 1.0 mg of dexamethasone sodium phosphate or dexamethasone acetate may be contraindicated. (includes all leads which contain these steroids) Epicardial leads should not be used on patients with a heavily infracted or fibrotic myocardium. The SelectSecure Model 3830 Lead is also contraindicated for the following: patients for whom a single dose of 40.µg of beclomethasone dipropionate may be contraindicated. patients with obstructed or inadequate vasculature for intravenous catheterization.

50 Brief Statement: Medtronic Leads (continued)
Warnings/Precautions People with metal implants such as pacemakers, implantable cardioverter defibrillators (ICDs), and accompanying leads should not receive diathermy treatment. The interaction between the implant and diathermy can cause tissue damage, fibrillation, or damage to the device components, which could result in serious injury, loss of therapy, or the need to reprogram or replace the device. For the SelectSecure Model 3830 lead, total patient exposure to beclomethasone 17,21-dipropionate should be considered when implanting multiple leads. No drug interactions with inhaled beclomethasone 17,21-dipropionate have been described. Drug interactions of beclomethasone 17,21-dipropionate with the Model 3830 lead have not been studied. Potential Complications Potential complications include, but are not limited to, valve damage, fibrillation and other arrhythmias, thrombosis, thrombotic and air embolism, cardiac perforation, heart wall rupture, cardiac tamponade, muscle or nerve stimulation, pericardial rub, infection, myocardial irritability, and pneumothorax. Other potential complications related to the lead may include lead dislodgement, lead conductor fracture, insulation failure, threshold elevation or exit block. See specific device manual for detailed information regarding the implant procedure, indications, contraindications, warnings, precautions, and potential complications/adverse events. For further information, please call Medtronic at and/or consult Medtronic’s website at Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.

51 Disclosure NOTE: This presentation is provided for general educational purposes only and should not be considered the exclusive source for this type of information. At all times, it is the professional responsibility of the practitioner to exercise independent clinical judgment in a particular situation.


Download ppt "Pacemaker Patient Follow-up Module 8"

Similar presentations


Ads by Google