Tungsten Cable Failure Analysis Brad James Ph.D. P.E.
2
Tungsten Cable Failure Analysis Background Medical device company uses tungsten cables as part of a surgical apparatus Non-sterilized cables withstand hundreds of durability test cycles (consists of cyclic and static loading) Sterilized cables fail after a few durability test cycles Tensile testing shows no real difference between sterilized and non-sterilized cables
3
Tungsten Cable Failure Analysis
4
SterilizedNon-Sterilized
5
Tungsten Cable Failure Analysis Sterilized Cable After Durability Testing
6
Tungsten Cable Failure Analysis Sterilized Cable After Durability Testing
7
Tungsten Cable Failure Analysis Cables After Tensile Testing: Sterilized Non-Sterilized
8
Tungsten Cable Failure Analysis Hydrogen content tests (Leco Interstitial) Non-sterilized cable: ~20 ppm hydrogen Sterilized cable: ~ 60 ppm hydrogen
9
Tungsten Cable Failure Analysis Conclusions Classic case of hydrogen embrittlement! (tungsten is known to be susceptible) Corrosion during sterilization is the source of hydrogen Embrittlement is stress/time dependent Tensile testing not long enough to allow diffusion of hydrogen to high stress locations Static loads during durability allowed hydrogen to diffuse to higher stress locations – resulting in fracture
Similar presentations
© 2018 SlidePlayer.com Inc.
All rights reserved.
Product mix ppt on nestle water Ppt on neural control and coordination Ppt on world environment day 2016 Ppt on political parties and electoral process in usa Ppt on indian festivals in hindi Download ppt on minimum wages act 1948 Ppt on library management system download Ppt on ntfs file system Ppt on e-waste free download Ppt on coalition government in uk