1 A Study on the Determination of Scan Speed in Whole Body Bone Scan Applying Oncoflash Gwang Gil Yang Department of Nuclear Medicine, Asan Medical Center, Seoul, Korea

2 Introduction Whole Body Bone Scan Radiopharmaceutical Injection Wholebody Scan (required about 15mins) 3~4 hrs delay

3 Purpose Save of scan time Maintenance of Image quality Oncoflash Application Proper Criteria

4 Materials and Method Quantitative evaluation Classified by the total counts - under 800K, over 800K, 900K, 1,000K, 1,500K, and 2,000K. Duration / Subject - July, 2008 / 329 patients Correlation evaluation Correlation analysis - Total counts (30cm/min) and PPM counts Duration / Subject – August, 2008 / 152 patients FWHM evaluation FWHM comparison with 4-Quadrant bar phantom - before and after applying the Oncoflash

5 Materials and Method Equipment / Collimator SIEMENS Symbia T2, E.CAM Signature LEHR Collimator Analysis program Syngo MI Application : Series FWHM Proc SPSS 12.0 Matrix size : 256 x 1024 / Window center : 140kev Window width : ±15% Acquisition Parameter

6 Materials and Method Definition of PPM Counts (Kilo Counts/sec, kcts/s)

7 Materials and Method FWHM measurement

8 Results Scan Speed : 30 cm/min Scan Speed : 15 cm/min Not applied Applied Not applied Applied Quantitative evaluation Total Counts - under 1,000K – : coarse particles and increased noises. (* Same Patient)

9 Results Total Counts (K)Geometric Mean ± SD (K)PersonRatio ~ 800762 ± 3382.4% 17.6 % 800 ~ 900851 ± 27206.1 % 900 ~ 1000953 ± 29309.1 % 1000 ~ 15001228 ± 14019759.9 % 82.4 % 1500 ~ 20001699 ± 1486620.1 % 2000 ~2294 ± 27182.4 % Total329100.0 % Quantitative evaluation Total Counts - under 1,000K – : coarse particles and increased noises. The Percentage according to Total Counts

10 Results FWHM Value after applying the Oncoflash : PPM counts of under 3.6K, were higher. : Whereas over 3.6K, were lower. FWHM evaluation (kcts/s) FWHM Comparison of FWHM Value between Applied and Not-applied Oncoflash

11 Results PPM counts (KiloCounts/sec) Total Counts ± SD (K)PersonRatio 2.5 ~ 3.0K965±173K 1617.8% 3.1 ~ 3.5K1084±154K 3822.4% 3.6 ~ 4.0K1242±186K 2218.4% 4.1 ~ 4.5K1359±170K 3117.1% 4.6 ~ 5.0K1405±184K 2010.5% 5.1 ~ 6.0K1640±376K 1810.5% 6.1 ~ 7.0K1771±324K 41.3% 7.1K ~1972±385K 32.0% Total 152100.0% Correlations between PPM counts and Total counts : strong correlation (r=.775, p<.01). Correlation evaluation Correlation between PPM Counts and Total Counts

12 Results (kcts/s) Correlations between PPM counts and Total counts : strong correlation (r=.775, p<.01). Correlation evaluation Total Counts Correlation between PPM Counts and Total Counts

13 Results PPM countWholebody count PPM count Pearson correlation coefficient 1.775** significance probability.000 Total sum of square, cross product 147/05135666.745 covariance.974236.204 N152 Wholebody count Pearson correlation coefficient.775**1 significance probability.000 Total sum of square, cross product 35666.7451438036 covariance236.20495291.627 N152 Correlations between PPM counts and Total counts : strong correlation (r=.775, p<.01). Correlation evaluation Pearson Correlation Coefficient between PPM Counts and Total Counts

14 Conclusion PPM counts - over 3.6K - & Total counts - over 1,000K - - The image applying the Oncoflash was similar in quality to raw image (not applying Oncoflash). - Reduce scan time without any damage on the image quality. Total counts - under 1,000K – - Oncoflash is applied, Image quality were decreased. - Recommended to perform the re-image in the scan speed of 15cm/min.