1 Qing Liang, PhD Medical Physicist Mercy Health System, Janesville, WI
Dose Measurement Qing Liang, PhD Medical Physicist Mercy Health System, Janesville, WI

2 Learning Objectives Learn about ionization calibration chain
Understand TG51 on photon beam and electron beam dosimetry Overview of TRS398

3 References AAPM TG 51: Med Phys 26(9), 1999
AAPM TG 51 Addendum 41(4),2014 IAEA TRS 398 Khan, The Physics of Radiation Therapy AAPM virtual library (website)

4 Part 1: Calibration Traceability
(ionization chamber)

5 Calibration Chain 2009AAPM summer school

6 What is a PSDL? 2009AAPM summer school

7 Primary Dosimetry Standard
2009AAPM summer school

8 Secondary Standard Dosimetry Lab
2009AAPM summer school

9 NIST National Institute of Standard and Technology
Dose measured by calorimeter, 60Co source NIST calibrated chamber :Absorbed dose in water rate (Gy/s) I: ionization current (C/s)

10 ADCL 2009AAPM summer school

11 Clinical calibration factor
NIST ADCL ADCL ADCL Clinic Clinic Clinic Clinic Clinic Clinic

12 Part 2: AAPM TG51

13 AAPM TG51

14 AAPM TG51

15 TG51 Overview: photon 2009AAPM summer school

16 TG51 Overview: electron 2009AAPM summer school

17 TG51 Overview 2009AAPM summer school

18 TG51 Overview 2009AAPM summer school

19 TG51 Overview 2009AAPM summer school

20 TG51 Phantom 2009AAPM summer school

21 TG51 Chamber 2009AAPM summer school

22 TG51 Effective point of Measurement
2009AAPM summer school

23 TG51 Effective point of Measurement
2009AAPM summer school

24 TG51 Chamber Reading Pion: recombination correction factor
Two voltage method (<1.05)

25 TG51 Chamber Reading PTP: temperature pressure correction
Correct for density compare to standard environmental condition: 22oC, kPa (760 mm Hg)

26 TG51 Chamber Reading Pelec: electrometer calibration factor
Provided by ADCL Unit: C/C or C/rdg If calibrated with chamber, Pelec=1 If calibrated separately, Pelec applied

27 TG51 Chamber Reading Ppol: polarity correction factor
Depends on chamber design, cable position and beam quality ( )

28 TG51 Beam Quality (P) 2009AAPM summer school

29 TG51 Beam Quality (P) 2009AAPM summer school

30 TG51 Beam Quality (P) 2009AAPM summer school

31 TG51 Beam Quality (P) OR 2009AAPM summer school

32 TG51 Beam Quality (P) 2009AAPM summer school

33 TG51 Photon Dosimetry 2009AAPM summer school

34 TG51 Photon Dosimetry 2009AAPM summer school

35 TG51 Photon Summery 2009AAPM summer school

36 TG51 Worksheet (P)

37 TG51 Beam Quality (E) 2009AAPM summer school

38 TG51 Beam Quality (E) 2009AAPM summer school

39 TG51 Beam Quality (E) 2009AAPM summer school

40 TG51 Beam Quality (E) 2009AAPM summer school

41 TG51 Beam Quality (E) 2009AAPM summer school

42 TG51 Beam Quality (E) 2009AAPM summer school

43 TG51 Beam Quality (E) 2009AAPM summer school

44 TG51 Beam Quality (E) 2009AAPM summer school

45 TG51 Beam Quality (E) Kecal 2009AAPM summer school

46 TG51 Electron Dosimetry 2009AAPM summer school

47 TG51 Beam Quality (E) 2009AAPM summer school

48 TG51 Electron Dosimetry 2009AAPM summer school

49 TG51 Electron Dosimetry 2009AAPM summer school

50 TG51 Electron Dosimetry 2009AAPM summer school

51 TG51 Electron Summery 2009AAPM summer school

52 TG51 Worksheet (E, cyl)

53 TG51 Worksheet (E, pp)

54 TG51 Summery 2009AAPM summer school

55 TG51 Addendum

56 TG51 Addendum Covers No lead foil to be used anymore (use equation)

57 Part 3: IAEA TRS 398

58 IAEA TRS 398

59 TRS 398 vs. TG 51 Basic equations are the same between TRS398 and TG51, expect notations Main difference is the beam quality parameter kQ: TPR20,10 in TRS 398, vs. %dd(10)x in TG51

60 TRS 398 Setup: SAD Source to chamber distance: 100 cm
Field size: 10x10 cm2 Depth: 20 cm, 10 cm

61 Thank you!