1 Good laboratory practices with waived test systems
January 2002 Washington State Department of Health Office of Laboratory Quality Assurance (206)

2 What is a “waived” test? Under CLIA, tests are categorized by complexity Waived, PPMP, Moderate, High Waived tests are: Simple lab examinations or procedures Cleared by FDA for home use Negligible likelihood of erroneous results No reasonable risk of harm if performed incorrectly

3 What does “waived” mean?
Waived from most requirements established for tests of higher complexity Site inspections are not routinely performed Proficiency testing is not required Personnel qualifications are not established

4 What do the Washington State regulations say for waived testing?
Obtain a Medical Test Site license Tell us which waived tests you perform Follow manufacturer’s instructions for performance of the test

5 Waived tests are simple, but …
Any test can produce erroneous results if not performed properly Any test worth running should be associated with good laboratory practices What are good laboratory practices? What are good risk management practices?

6 Have a current product insert
Make sure it states “CLIA waived” For some tests (defined in regulations) – all are waived (i.e., fecal occult blood) For some tests – only certain specific test systems are waived (i.e., Strep antigen) For some test kits – test may be waived or moderate complexity, depending on specimen used (i.e., mononucleosis – whole blood is waived, serum or plasma is moderate complexity

7 Focus on these sections
Intended use Product storage Precautions/Warnings Patient preparation Specimen collection Test procedure Procedural notes Quality control Results/Interpretation Limitations Expected results

8 In the following slides, we give examples (in quotations) from actual product inserts to show the kinds of information you will find in the various sections of the product insert

9 Intended use Describes what is actually being measured
“measures percent concentration of HgbA1C in blood” Qualitative versus quantitative “qualitative detection of Group A Strep antigen directly from throat swab” Type of specimen “not recommended for use with gastric specimens”

10 Product storage Kits, reagents, test devices
“store refrigerated at 2 – 8 degrees C” “keep 3 months at room temperature” “cassettes must be stored in sealed foil pouches” “store out of sunlight” “strips must be kept in bottle with cap tightly closed”

11 Precautions / Warnings
Proper handling of kits, reagents, test devices “do not use past expiration” “do not mix components of different lots or kits” “should not be interpreted by individuals with blue color deficiency (color blindness)” “do not interchange caps on reagents”

12 Patient preparation Examples: CLO test Occult blood
“discontinue use of antibiotics and bismuth preparations 3 weeks before biopsy” Occult blood “for 7 days … avoid non-steroidal anti-inflammatory drugs, for 3 days … avoid vitamin C in excess of 250 mg a day, … avoid red meats …”

13 Specimen collection, handling
Acceptable types of specimens “finger stick or venipuncture” “can be stored at room temperature for 4 hours and up to 72 hours if refrigerated” “do not use swabs that have cotton tips or wooden shafts” Acceptable anticoagulants “acceptable anticoagulants are EDTA, heparin, citrate, fluoride”

14 Test procedure/Directions/Instructions
Follow exactly Don’t modify Adhere to timing

15 Procedural notes “do not open foil pouch until ready to test … avoid cross contamination … read results within 20 minutes” “directions must be followed exactly, accurate timing is essential” “allow specimen and test devices to warm to room temperature before use”

16 Quality control The types of controls to be tested vary with the specific test system used External, Internal, Electronic Test controls according to the manufacturer’s instructions Read the entire Quality Control section carefully Assure that you get the expected results for the controls tested

17 Quality control External controls
Reference solutions or materials (i.e., swabs) Added to test device like the patient sample May be included with the test kit or you may need to purchase separately Compare your control results to the expected ranges or values printed on the control vials or in the control product insert

18 Quality control Internal (built-in, procedural) controls
Built into test reagent devices to ensure that reagents are active, reagents & samples are added correctly, test performs according to specifications Common with qualitative tests (Strep antigen, pregnancy, H. pylori, mononucleosis) Procedural controls typically include the appearance of a colored dot, line or bar in a control region and/or an expected appearance of the device background

19 Quality control Electronic controls Inert, reusable devices
(test strips, cartridges, cassettes) Used to check instrument performance specifications Available for use with some quantitative test systems (hemoglobin, cholesterol, A1C, prothrombin time) Compare the control results with the expected values

20 Results / Interpretation
Positive Negative Invalid Reportable range of method “patient values are linear from 2.5% to 14.0%” “linear up to 23.5 g/dl”

21 Limitations of procedure
Causes of false positives, false negatives “patients must have a total hemoglobin between 7 and 24 g/dl” “measuring range is 100 to 500”

22 Expected results, values
“the following values are considered normal…” “normal range is 4.3% - 5.7%” “95% of prothrombin time values range from 10.6 to 13.4 seconds” “for the purposes of calculating the INR, normal is defined as 12.0 seconds” “will detect 20 mIU/ml of HCG”

23 Good laboratory practices
Test controls - Follow manufacturer instructions “positive and negative controls should be tested with each new lot or shipment of test materials” “Daily Requirements: Two levels of electronic quality control or two levels of liquid controls must be tested…” “a positive and negative control must be tested when opening a new test kit and with each change in operator within the test kit” “check the calibration daily by using the control cuvette”

24 Good laboratory practices
At minimum, test external controls with each new lot of kits, reagents, testing devices To detect problems during shipment Observe internal (built-in, procedural) controls with each patient test To assure proper test performance, reagent integrity Test electronic controls periodically According to manufacturer’s instructions & frequency

25 Good laboratory practices
Correlate test results with patient presentation, history, diagnosis Participate in a proficiency testing (PT) program Many PT companies offer 2 sample programs Split sample program Compare your results with reference laboratory

26 Training Have new employees read the entire product insert
Not just the Quick Reference Card Keep a record of training Have each employee demonstrate competency Initially Periodically

27 Keep a simple log of results
By keeping results of the following in chronological order, you can detect potential problems: New lots of reagents, kits, testing devices Quality control results Patient test results

28 Report test system problems
To manufacturer To Food & Drug Administration’s MedWatch Call FDA-1088

29 Questions? Call the Office of Lab Quality Assurance: