Presentation is loading. Please wait.

Presentation is loading. Please wait.

Biopotential Electrodes

Similar presentations


Presentation on theme: "Biopotential Electrodes"— Presentation transcript:

1 Biopotential Electrodes

2 Introduction Electrical Contact point Transducer
Biopotential electrodes Metal (Al, Cu, Fe, Ag,…..) Non-metal

3 Electrode – Electrolyte Interface
Electrode Electrolyte (neutral charge) M M+ and A- in solution Current flow A- M+ M e- M M+ e- A- e- M+ M  M+ + e- A-  A + e- M+ : Cation A- : Anion e- : electron

4 Metal Electrolyte Interface
To sense a signal a current I must flow !

5 The Interface Problem To sense a signal a current I must flow !
But no electron e- is passing the interface!

6 Metal Cation leaving into the electrolyte No current What’s going on?

7 Metal Cation: leaving into the electrolyte
No current One atom M out of the metal is oxidized to form one cation M+ and giving off one free electron e- to the metal.

8 Metal cation: joining the metal
No current What’s going on?

9 Metal Cation: joining the metal
No current One cation M+ out of the electrolyte becomes one neutral atom M taking off one free electron from the metal.

10 Half-cell Voltage No current

11 Half-cell Voltage No current metal: Li Al Fe Pb H Ag/AgCl Cu Ag Pt Au
Vh / Volt negativ positiv

12 Electrode Double Layer
No current ? ? ?

13 Electrode Double Layer
No current ? ?

14 Electrode Double Layer
No current ?

15 Electrode Double Layer
No current Oxidation or reduction reactions at the electrode-electrolyte interface lead to a double-charge layer

16 Contact (Half Cell) Potential
Depends on: The metal, Concentration of ions in solution and Temperature. Half cell potential cannot be measured without a second electrode. The half cell potential of the standard hydrogen electrode has been arbitrarily set to zero.

17 Measuring Half Cell Potential
Note: Electrode material is metal + salt or polymer selective membrane

18 Half Cell Potential (Vh)
Iron mV Lead mV Copper mV Platinum mV Compare to electrophysiological Signals ??? Two Similar electrodes ??? (Ag/Agcl  5 mV and steel 100mV)

19 Difference bet. observed and zero-current half cell potentials
Polarization If there is a current between the electrode and electrolyte, the observed half cell potential is often altered due to polarization. Overvoltage Difference bet. observed and zero-current half cell potentials Resistance Current changes resistance of electrolyte and thus, a voltage drop results. Concentration Changes in distribution of ions at the electrode- electrolyte interface Activation The activation energy barrier depends on the direction of current and determines kinetics

20 Polarizable and Non-Polarizable Electrodes
Perfectly Polarizable Electrodes No actual charge crosses the electrode-electrolyte interface when a current is applied. (e.g Platinum electrode) Perfectly Non-Polarizable Electrode Current passes freely across the electrode-electrolyte interface. These electrodes see no overpotentials. (e.g. Ag/AgCl Electrode) Example: Ag-AgCl is used in recording while Pt is use in stimulation

21 Ag/AgCl Electrode Fabrication of Ag/AgCl electrodes
Electrolytic deposition of AgCl Sintered AgCl: process forming pellet electrodes

22 Electrolysis Process


Download ppt "Biopotential Electrodes"