Presentation is loading. Please wait.

Presentation is loading. Please wait.

Customized LASIK & PRK will dominate in next few years

Similar presentations


Presentation on theme: "Customized LASIK & PRK will dominate in next few years"— Presentation transcript:

1

2 Customized LASIK & PRK will dominate in next few years
Customized Corneal Ablation Customized LASIK & PRK will dominate in next few years Speedy recovery, good quality of vision satisfactory outcome Disadvantage of conventional refractive surgery in some patients: * Increase in HOA * Reduction in visual quality ٤٩

3 Wavefront Customized Visual Correction Ocular wavefront sensing:
Will be increasingly employed Will become routine in vision assessment Wavefront customization is employed to optimize any refractive surgery procedure ٥١

4 Wavefront Customized Visual Correction (cont)
Future wavefront customized refractive procedures Implantation of optimized IOL’s e.g Technis aspheric lens Customized IOL’s preinsersion, customized phakic IOL’s

5

6 Types of customization
Two main methods are available in refractive surgery: Topography guided ablation Wavefront guided ablation

7 According to recent statistics, approximately 55% of North American refractive surgeons have wavefront analyzers in their practice and routinely perform wavefront-guided ablations.

8 Definition of Wavefront

9

10

11

12

13

14

15

16 Wavefront-Guided (Customized) Excimer Laser Refractive Surgery
Definition: Wavefront-guided custom ablation is used to correct higher-order aberrations, in addition to spherocylindrical correction.

17 Incidence in General population
10-15% have significant higher-order aberration

18

19 Hard contact lenses Corrects higher order aberrations resulting from the cornea.

20

21 How to measure the higher order optical aberration?
By using a Wavefront analysis system called aberrometry

22 Pupil and Wavefront Larger diameter leads to a larger Wavefront error

23

24

25 Evaluation of vision quality
Compromised more at dim light during night and represented by: a) Double vision b) Ghosting c) Glare d) Halos e) Starbursts d) Reduced contrast sensitivity

26 Quantification of magnitude of the aberration
1)Usually by RMS wavefront error 2)It gives a rough estimate

27 Types of aberrations Low Order Aberration (LOA) : 1st & 2nd order aberration High Order Aberration (HOA) : 3rd to 6th or 10th

28 How to evaluate the quantity of optical system
Root Mean Square ( RMS ) The RMS is SD of height (depth) of wavefront relative to the reference at all the point in the wavefront. RMS is very useful measure of optical quality. Peak-To-Valley (PV) Distance from the highest to lowest point on the deformed wavefront relative to the reference wavefront.

29 RMS1 is qualitative presentation of 1st order aberration, RMS2 is due to 2nd order and etc.
RMSh represents the total RMS of HOA (including 3rd to 6th OA) RMSg represents the total RMS of HOA and LOA Total RMS increases with aging LASIK increases RMS especially in subjects with larger pupils RMSh > 0.3 and > 20% of total RMS is clinically significant and indicative for customized ablation

30 Wavefront measuring devices
Hartmann-shack style devices are the most common used

31

32 Hartmann-shack style 1)A narrow beam of light is projected on to the retina and the light reflected from the fovea passes through the lens and the cornea and exists the eye

33 Customization can be based on corneal topography or wavefront measurements.
Corneal topography guided ablation has been attempted on patients with regular and irregular astigmatism, decentered ablations, and central islands. The irregular astigmatism group is more challenging and may benefit most from corneal topography guided ablation as the systems become more refined.

34

35 Hartmann-shack aberrometer
A low intensity laser beam is directed to retinal surface Light rays from this laser spot are reflected back to the front of the eye

36

37 Laser spot on retinal surface
light rays are reflected back to the front of the eye Lens arrey focuses these light rays Photodetector (charged coupled device,CCD) Dots of light

38

39

40

41 Considering refractive lens at corneal surface
Real eye with Aberrations Considering refractive lens at corneal surface Excimer etching by customized ablation Converting wavefront to flat wavefront Perfect focus on fovea Ideal eye ١٢

42

43

44

45

46

47

48

49 Zero order (no order)=axial symmetry, flat
Zernike’s Polynomials Zero order (no order)=axial symmetry, flat wavefront First order = linear aberration, tilting around a horizontal (x) or vertical axis (y) Second order = focus shift, spherical defocus or astigmatism Third order = corresponding to coma, triangular astigmatism Fourth order = spherical aberration, complex patterns Fifth-10th order = Irregular aberrations, important when pupil is wide dilated. 10/14/2007 ٧ XVth Congress of Iranian Society of Ophthalmology October 2005

50

51

52

53

54

55

56

57

58

59

60 Wavefront guided customization Goals
Spherocylinder correction To correct the aberration that are induced by conventional laser vision correction and pre-existing aberration.

61 Requirment for wavefront
Efficient eye tracking Good registeration system for eye Small size laser spot Sufficient corneal bed thickness

62 Challenges to wavefront
Tear film abnormalities Miotic pupil : If severe may be difficult to measure and provide information beyond the 3mm optical zone and therefore require pharmacological dilation.

63

64 Optical effects of tear film disruption

65 Wavefront map with cycloplegic or mydriatic
1) Has significant difference in pre-op refractive error compared to subjective refraction

66 (Ablation profile algorithm)
1) After good aberration maps, Ablation profile algorithm is needed 2) Should be designed to make the post-op higher order aberrations as minimal as possible 3) Not only correcting the pre-op higher order aberration but also preventing the induction of new higher order aberrations.

67 Conversion to Ablation Profile:
Ablation profile is fundamentally inverse of wavefront error map Goal: Correction of refractive error and higher-order aberrations Pupil diameter at least 0.5mm larger than scotopic Limbal marking for cyclotorsion detection Wavefront measurement + corneal curvature + biomechanics ablation profile complex ٣٥

68

69 Emphasis on Registration

70 Registration Registration refers to the method by which a customized ablation pattern based on preoperative wavefront analysis from the aberrometer is superimposed in the same position on the eye during the laser treatment.

71 LADARWave and LADARVision systems:
Centration is established by the surgeon when the patient fixates on the target light. The instrument captures a video image of the eye, and the surgeon defines the pupillary ring (red) and limbus ring (yellow). These will be used by the LADARVision tracker to follow the eye during application of the customized ablation pattern.

72 Torsional alignment Torsional alignment refers to the need to compensate for the cyclotorsion of the eye that occurs when the patient lies down. The eye usually rotates less than 5 degrees, but in some patients, postural cyclotorsion can be as much as 10 to 12 degrees.

73 Misalignment A small misalignment in in the axis can have significant impact on the results of procedure A 5-7 degrees of cyclotorsion when changing from sitting position to supine position. 50% of the visual benefit of HOA is lost.

74

75 Bausch and lomb zyolptix system has
Diagnostic Treatment components for wavefront customized corneal ablation

76 Diagnostic part of zioptix systems (2172 laser)
In zy wave aberometer and ORBscan both machine are integrated in one workstation. It has Hartmann-shack sensor that measures up to 5th order zernike term. It measures sphere from +6.0 D to -12. D and 0 to 5 D of cylender with pupil diameter from 2.5 to 8.6 mm.

77 Pupil during exam Orbscan must be performed with an undilatd pupil
Wavefront meaurements are taken with a dilated pupil.

78

79 Types of beam Has a truncated gaussion 3 beam to achieve the benefits of both beam types. 1 2 3

80 Laser beam sizes in Tech217z
The 2mm beam treat the majority of the refractive errors. The 1mm beam used for more specific ablation pattern on the transition zone

81

82 Eye tracker The system has 120 Hz active eye tracker with a passive automatic shut-off system corrects for intraoperative eye movement in every dimention including cyclotortion and pupil shift

83 FDA investigation for outcomes for wavefront guided lasik in myopia for 3 separate laser platforms
LADAR vision Technolus 217 z for low to moderate myopia visx Conventional Wave front guided 72% ………….20/20 89% ……..20/20

84 Environmental Conditions
Control of environmental conditions during treatments is important. In previous U.S. FDA Multi-Center Clinical Trials, the room conditions were: Temperature ranged from 68ºF to 72ºF (20ºC to 22.2ºC) Relative humidity ranged from 40% to 45% Treatments performed at >24ºC were associated with less accurate outcomes Stability of temperature and humidity is important

85 Future of Customized Corneal
Ablation LASIK vs surface ablation Each microkeratome induces specific “flap only” aberration (flap size, thickness) Considering “flap aberrations” in total treatment calculations ٤٢

86 Show promising results with use of
Future of Customized Corneal Ablation … cont. Surface Ablations: Show promising results with use of immunomodulating agents Better control of cellular and biochemical reactions Introduction of new drugs to better regulate wound healing and refractive outcome Gene therapy for better control of post laser keratocyte activation and wound healing ٤٣

87

88 Presbyopia: Customized multifocal ablation
Aberrations may be induced when creating multifocality Potential loss of contrast sensitivity and quality of visual function Future results will be improved: * Wavefront mapping, sophisticated eye trackers and corneal registration * Preoperative simulation of postop condition ٤٤

89 Correction of HOA needs smaller spot delivery
Laser Delivery Refinements Correction of HOA needs smaller spot delivery requires mm spot size >5th order Smaller spot size needs faster and better eye trackers Smaller ablation depth per pulse provides ideal correction profile for higher orders ٤٥

90 The Human Hair (125 microns) Etched with an excimer laser
Courtesy of IBM

91 Effects of Laser Ablation
Thermal Non-thermal YAG CO2 Excimer

92

93

94 Wavefront measurement devices and consequently wavefront correction
Conclusion Wavefront measurement devices and consequently wavefront correction procedures are still in process of evolution Achievement of “supervision”: with advancement in current procedures will not be a dream in near future ٥٤

95

96 Thanks for your attention


Download ppt "Customized LASIK & PRK will dominate in next few years"

Similar presentations


Ads by Google